- Applicable for 96- / 384- / 1536-well plates
- All resin tip realizes relatively inexpensive running cost.
- Serial dispensation to 8, 12, 16 and 24 wells and serial dilution is available as well as 96 / 384ch dispensation.
- High throughput can be achieved with connecting Multi Stacker 400
Multi Works 600(96/384ch automated pipetting systems)
96 / 384ch automated pipetting system using disposable tips
Product features
Ultra small volume dispensation from 25 nano litter ( planned ) with disposable tips has come true.
Nano Litter Dispensing Head
Nano Litter Dispensing Head is planned to achieve dispensing volume from 25 to 1,000 nano litter.
Since it is also applicable for existing micro litter dispensing heads, it can dispense in very wide range, from 25 nano litter to 300 micro litter, with the same single equipment and high accuracy.
Special nano litter disposable tip
Special disposable tip for nano litter dispensation Using disposable tip prevents contamination problem.
All resin tip realizes relatively inexpensive running cost.
High work efficiency with good cost performance
Applications in various fields
Applicable for vast variety of applications like pre-screening and optimization in protein crystallization.
Automatic Tip Changer
It replaces tips from a tip rack on a stage automatically, inserting the nozzle holder under the pipetting head.
If you take tip changes automatic, you don't have to replace tip by hand.
Manual Tip Changer
Insert the tip holder at the bottom of the pipetting head and replace it. When the manual tip replacement method is adopted, no tip rack is required, so there is no space.
Pipetting Head
The system has high maintainability because users can change the pipetting heads very easily.
Dispensing Stage
The system has 8 stages and each can accommodate SBS standard vessels. There is no need for replacing tips and plates while transferring liquid in 4 96-well plates to 1 384-well plate through the operation.
Dispensing Mode
Serial dispensation to 8, 12, 16 and 24 wells and serial dilution is available as wel as 96 / 384ch dispensation.
Software
User friendly and easy to understand control software lets everybody write programs.
Model list
Dispensing head | 96ch/1ul |
---|---|
Dispensing volume | 0.025〜1ul |
Dispensing | 100nl CV≦5% 500nl CV≦3% |
Dispensing head | 96ch/50ul |
---|---|
Dispensing volume | 1〜40ul |
Dispensing | 2ul CV≦4% 30ul CV≦2% |
Dispensing head | 96ch/300ul |
---|---|
Dispensing volume | 5〜300ul |
Dispensing | 5ul CV≦3% 100ul CV≦1% |
Dispensing head | 384ch/1ul |
---|---|
Dispensing volume | 0.025〜1ul |
Dispensing | 100nl CV≦5% 500nl CV≦3% |
Dispensing head | 384ch/50ul |
---|---|
Dispensing volume | 1〜40ul |
Dispensing | 2ul CV≦4% 30ul CV≦2% |
Specifications
Dispensing system | Disposable tip with positive displacement Air displacement disposable tip replacement |
---|---|
Tip change | Automatic Chip changer or Manual |
Dispensing Mode | 996 / 384ch, 8 / 12 / 16 / 24ch serial dispensation, 8 / 12 / 16 / 24ch serial dilution |
Work area | 8 stages |
Control | Windows PC |
Interface | RS232C |
Size | 560 mm (W) X 504mm mm (D) X 625mm mm (H) Stage work space : 295mm (right side) / 300mm (left side) Retractable keyboards : 127 mm |
Weight | approx. 86kg |
Power | AC100 - 240V 50/60Hz (Power consumption : 130W ) |
Options | Plate stacker, tip rack stacker, plate handling arm Automatic supplier for reservoir, tip flusher |
Option
Multi Stacker 400 can be set both right and left side of Multi works 600 each. ( up to 2 stackers )
Configuration
Equipment | Multi Works 600 (10) AC power cord (1) |
---|---|
Document | Warranty (1) Manual (1) |
Recently, minimizing quantity of regent gets very popular to cut running cost. Especially testing in nano litter range that could be performed less budget, reduces both cost of regent and use of valuable specimen.
However, if minimizing quantity makes dispensing accuracy worse, that is putting the cart before the horse.
Multi Works 600 is the automated pipetting system that can respond to user's cost reduction needs in higher level while it is improving reliability in test results under minimizing dispensing volume.